Overview
Biological
Crystallography
  BL13B1
  BL13C1
  BL12B2
X-ray Absorption
Spectroscopy
Small Angle X-ray
Scattering
Membrane
Scattering
Fiber/Powder
Diffraction
Microscopy and
Imaging
Bio-Sensor Fabrication
and Characterization
Synchrotron Radiation
Circular Dichroism
BL13C1
    Contact Information ~
    If you have any technical concerns, please contact with us!
    BL13C1 Spokesperson:
      Yuch-Cheng Jean (ycjean@nsrrc.org.tw, +886-3-578-0281 Ext: 7321)
    BL13C1 Beamline Manager:
      Cheng-Hung Chiang (jjh@nsrrc.org.tw, +886-3-578-0281 Ext: 1133)

    Technical Information of BL13C1 ~
    Beamline Name
    NSRRC_BL13C1
    Beamline Website http://bionsrrc.nsrrc.org.tw
    Owner/Operator NRPGM-SPXF/NSRRC-PX Group
    Operational (general use) Yes, Since Sep. 21th, 2006.
    %Time available (general use) 65%
    Beamtime Schedule http://portal.nsrrc.org.tw/schedule/lsch.php
    Experiments HR-SAD, & Monochromatic
    Source 28 Poles, 6 cm Period Length, 3.5 Tesla Superconducting Wiggler
    Ring Current(mA) 300 mA
    Average Beam Lifetime Infinite Under NSRRC Top Up Conditions
    Energy Range 12.4, 12.7, & 13.6 keV (1 Å, 0.976 Å, & 0.915 Å)
    Energy Resolution 8 x 10E(-4)
    Spot Size FWHM: 0.68 mm (H) x 0.15 mm (V)
    Optics Vertically Focusing Mirror, Horizontally Focusing Single Crystal Si(111) Bent Monochromator
    Flux 4 x 10E(10)
    Flux Detailed Flux Measured Through a 0.2 mm Round Aperture @ 12.7 keV
    Wavelength Min (Å) 0.915
    Wavelength Max (Å) 1
    Monochromator Horizontally Focusing Single Crystal Monochromator
    Crystal Type Si(111)
    Mirrors Rh Coated
    Mirror Cutoff 15 keV
    Special Capabilities Automated Sample Mounting
    Goniometer Huber Single Axis Goniometer
    Pin Length Manual Sample Mounting: No Restrictions;
    Automated Sample Mounting: 14 – 18 mm
    Restrictions Omega +/- 720 degree
    Xtal-detector Distance 100 - 810 mm
    Detector ADSC Quantum-210 CCD Area Detector
    Detector Details Inter-Frame Total Dead Time: 5s
    2theta Range Vertical Offset + 230 mm, -16 mm; Horizontal Offset +/- 80mm
    Cryo Capability Rigaku/MSC X-Stream2000, ~110 K @ Sample
    Biohazard Contact in Advance
    Software BluIce Control Software, HKL/HKL2000, Strategy, CHOOCH
    Hardware 3 2-CPU SGI Octane2s, 1 4-CPU SGI O350 Server, 1 Shared 12-CPU SGI Origin3800 Server, Shared 2.0 TB SGI Fiber Channel Raid System
    Backup Devices USB2.0 / IEEE1394 Portable Disk

    Photos of BL13C1 ~

    BL13C1 Mono Experimental Station

    Close View of Sample Surrounding

    Crystal Handling Area

    Experimental Control Area