Overview
Biological
Crystallography
X-ray Absorption
Spectroscopy
Small Angle X-ray
Scattering
Membrane
Scattering
Fiber/Powder
Diffraction
  BL01C2
Microscopy and
Imaging
Bio-Sensor Fabrication
and Characterization
BL01C2
    Contact Information ~
    If you have any technical concerns, please contact with us!
    BL01C2 Spokesperson:
      Hwo-Shuenn Sheu (hsheu@nsrrc.org.tw, +886-3-578-0281 Ext: 7122)
    BL01C2 Beamline Manager:
      Wei-Tsung Chuang (weitsung@nsrrc.org.tw, +886-3-578-0281 Ext: 7119)

    Technical Information of BL01C2 ~
    Beamline Name
    NSRRC_BL01C2
    Owner/Operator NSRRC/NSRRC-Powder Diffraction Group
    Operational (general use) Yes, Since Feb. 8th, 2005.
    %Time available (general use) 50%
    Beamtime Schedule http://portal.nsrrc.org.tw/schedule/lsch.php
    Experiments Monochromatic
    Source 5 Tesla Superconducting Wavelength Shifter
    Ring Current(mA) 300 mA
    Average Beam Lifetime Infinite Under NSRRC Top Up Conditions
    Energy Range 12 ~ 33 keV (1.03 Å ~ 0.37 Å)
    Energy Resolution 2.2 x 10E(-4)
    Spot Size FWHM: 1 mm (H) x 1 mm (V)
    Optics Vertically Collimating Premirror, Fixed-Exit Double Crystal Si(111) Monochromator, Toroidal Focusing Mirror
    Flux 1.4 x 10E(10)
    Flux Detailed Flux Measured Through a 0.2 mm Round Aperture Through Whole Spectra
    Wavelength Min (Å) 0.37
    Wavelength Max (Å) 1.03
    Monochromator Fixed-Exit Double Crystal Monochromator
    Crystal Type Si(111)
    Mirrors Rt Coated
    Mirror Cutoff 35 keV
    Goniometer Huber Single Axis Goniometer 410
    Pin Length No Restrictions
    Xtal-detector Distance 120 – 800 mm
    Detector Mar345 Imaging Plate Area Detector
    2theta Range Vertical Offset + 300 mm, - 300 mm; Horizontal Offset + 100 mm
    Cryo Capability ARS 4K – 350K
    High temperature Capability 300K – 1000K
    Software Mar345 Control Software, Fit2D, WinPlotr, GSAS, Spec
    Hardware Unix PC Control
    Backup Devices Windows PC, Data Analysis, Backup

    Photos of BL01C2 ~

    BL01C2 Experimental Station

    Close View of Sample Surrounding

    Fuji Bas 2500 IP Scanner

    Experimental Control Area